BDER-KI-11K

Silicon detection block with thermoelectric cooling BDER-KI-11K.

It is designed to convert the energy of X-ray and low-energy gamma radiation quanta into electrical signals proportional in amplitude and amplify them for subsequent recording by spectrometric equipment.

--11_3_copy

MAIN ADVANTAGES:

  • high energy resolution;
  • excellent background characteristics;
  • small dimensions and weight;
  • does not require liquid nitrogen;
  • arbitrary spatial orientation.

Application areas: nuclear energy, geology, metallurgy, environmental control systems, recycling, customs control and forensics.

Can be used as part of X-ray fluorescence analyzers for rapid determination of the elemental composition of a substance.

The detection unit is designed to operate in conditions corresponding to the UHL category 5.1 GOST 15150-69.

MAIN TECHNICAL CHARACTERISTICS:

No.

Name of the characteristic

Meaning

1

The area of the sensitive surface of the detector

7 mm2 and 12 mm2

2

Thickness of the sensitive area of the detector

300 microns

3

Energy resolution of 5.9 keV with a shaping time constant of 10 µs for the detector:

- with an area of 7 mm2

- with an area of 12 mm2

no more than 200 eV

no more than 250 eV

4

Energy resolution of 59.6 keV with a shaping time constant of 10 μs for the detector:

-with an area of 7 mm2

-with an area of 12 mm2

no more than 200 eV

no more than 250 eV

5

Ratio of the peak height of total absorption for energy 5.9 keV to the level of continuous amplitude distribution for energy 3.5 keV (peak/background)

not less than 800

6

Thickness of the entrance beryllium window:

- standard

- by special order

25 microns

12 µm and 8 µm

7

Conversion coefficient at load not less than 1 kOhm

not less than 1 mV/keV

8

Output impedance

75±1 Ohm

9

Output signal polarity

negative

10

Maximum detector bias voltage at current less than 1 µA

100V

11

Preamplifier supply voltage at maximum current 35 mA

12V

12

Maximum cooler current

0.7A at 1.6V

13

Weight

no more than 0.25 kg

14

Overall dimensions

15x61x34 mm

 

Fig. 1. Block connection diagram

Fig. 3 Efficiency of Si(Li) and Si-pin detectors depending on energy and thickness of Be-window

Fig. 2. Detection unit BDER-KI-11K